Page 7 Design and specifi cations subject to change without notice
Installation Manualwww.baumerprocess.com
Mounting of 3A Approved Products
Max. ± 85°
Max. ± 175°
5)
1)
2)
3)
4)
Installation of 3A approved products:
1) Use only a 3A approved counter part.
2) The inspection hole should be visible and drained.
3) Mount the instrument in a self drained position.
4) Level the inner surface of the pipe with the counter part.
5) Weldings should be grinded to Ra= 0.8
Refer to the data sheet “FlexBar Accessories” for O-rings, gaskets and
other accessories.